Description
The HiTemp140-PT is a high temperature, precision data logger with a bendable probe. Includes ISO17025 accredited calibration certificate.
Key Features:
- Submersible
- Available in Various Probe Lengths & Styles
- Measures Temperatures from -200 °C to +350 °C
The HiTemp140-PT is a submersible, high temperature data logger with a 12 or 24-inch bendable probe. This device is extremely versatile and can be tailored to the application with various probe lengths and styles available.
The HiTemp140-PT can withstand temperatures ranging from -40 °C to +140 °C (-40 °F to +284 °F), but is capable of measuring extended temperatures up to 350 °C (662 °F) (probe dependent). With a storage capacity of up to 32,256 time and date stamped readings, the HiTemp140 can be programmed for immediate start or delayed start up to 18 months in advance.
The HiTemp140-PT is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.
Body of data logger cannot exceed 140 °C (284 °F), but can extend to higher temperatures when used with MadgeTech’s thermal barrier options.
General
| Memory | 32,256 readings |
| Reading Rate | 1 reading every second up to 1 reading every 24 hours |
| Baud Rate | 125,000 |
| Battery Life | 2 years typical (1 minute reading rate)
Fast reading intervals and extreme temperatures will shorten the battery life substantially User Replaceable Battery |
| Material | 316 Stainless Steel, Radel |
| HiTemp140 Pig Tail Probe Dimensions | Body: 1.75 in x 0.97 in dia. (45 mm x 25 mm dia.)
HiTemp140-PT-1: Probe tip is 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.) HiTemp140-PT-5: Probe tip is 4.8 in x 0.125 in dia. with 1 in x 0.188 dia. handle (121 mm x 3.2 mm dia. with 25 mm x 4.8 mm dia. handle) HiTemp140-PT-12: Probe tip is 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.) |
| IP Rating | IP68 |
| Submersible | Yes |
| Operating Environment | -40 °C to +140 °C (-40 °F to +284 °F); 0 %RH to 100 %RH; 0.002 PSIA to 100 PSIA (Body of Logger Cannot Exceed 140 °C) |
| Required Interface Package | IFC400 |
| Approvals | CE |
Temperature
| Temperature Sensor | 100 Ω Platinum RTD |
| Temperature Resolution | 0.01 °C (0.02 °F) |
| Temperature Calibrated Accuracy | ±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F) ±0.3 °C/±0.54 °F (-20 °C to +19.99 °C/-4 °F to +67.98 °F) ±0.4 °C/±0.72 ºF (-40 °C to -20.01 °C/-40 °F to +-4.02 °F) |
| Probe Temperature Range | -200 °C to +350 °C (-328 °F to +662 °F) (Body of Logger Cannot Exceed 140 °C) |